Dosimetrical assessment of jaw tracking technique in volumetric modulated arc therapy for a sample of patients with lateralised targets

Published online by Cambridge University Press:  15 February 2023

Rogelio Manuel Diaz Moreno*
Affiliation:
Instituto Zunino, Córdoba, Argentina
Maria José Almada
Affiliation:
Instituto Zunino, Córdoba, Argentina
Albin Ariel García Andino
Affiliation:
Instituto Zunino, Córdoba, Argentina
Carlos Daniel Venencia
Affiliation:
Instituto Zunino, Córdoba, Argentina
*
Author for correspondence: Dr Rogelio Manuel Diaz Moreno, Instituto Zunino, 9 de Julio Street, 2015th, 10, Córdoba, X5003CQI, Argentina. E-mail: rdiaz@institutozunino.org

Abstract

Introduction:

In modulated radiotherapy treatments with the jaw tracking technique (JTT), the collimator jaws can dynamically follow the multileaf collimator apertures and reduce radiation leakage. This reduction protects normal tissue from unwanted doses. Previous research has highlighted the importance of defining which patients will benefit most from JTT. Besides, some authors have expressed their concerns about possible increases in monitor units (MUs). Treatments of patients with peripheral targets and isocentre located in the patient’s midline are of particular interest. The current work assessed the effect of JTT on these cases.

Methods:

JTT plans for thirty-two patients were compared to plans with the static jaws technique. The volumes of normal tissue receiving 5 Gy (V5), 10 Gy (V10) and 20 Gy (V20), mean dose (Dmean), target coverage parameters D95, D2% and Paddick’s conformity index (PCI) were compared. MUs were also registered for comparisons. The decrease in the jaws opening with JTT was correlated to the decrease in dose values in normal tissue.

Results:

Small decreases were observed in D95 and in D2% values, without statistical significance. A 5% average decrease in PCI values was noticed as well as significant decreases in V5, V10 and Dmean values, 9% on average. A 3% decrease in V20 was also observed. The number of MUs decreased by 2%. A significant correlation was found between the reduction of the secondary collimation opening areas and the dose delivered to normal tissue.

Conclusions:

JTT technique improved normal tissue protection in volumetric modulated arc therapy treatments for the patients included in the present study.

Type
Original Article
Copyright
© The Author(s), 2023. Published by Cambridge University Press

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Mohan, R, Jayesh, K, Joshi, R C et al. Dosimetric evaluation of 120-leaf multileaf collimator in a Varian linear accelerator with 6-MV and 18-MV photon beams. J Med Phys 2008; 33 (3): 114118.CrossRefGoogle Scholar
Glide-Hurst, C, Bellon, M, Foster, R et al. Commissioning of the Varian TrueBeam linear accelerator: a multi-institutional study. Med Phys 2013; 40 (3): 031719.CrossRefGoogle ScholarPubMed
Schmidhalter, D, Fix, M K, Niederer, P et al. Leaf transmission reduction using moving jaws for dynamic MLC IMRT. Med Phys 2007; 34 (9): 36743687.CrossRefGoogle ScholarPubMed
Joy, S, Starkschall, G, Kry, S et al. Dosimetric effects of jaw tracking in step-and-shoot intensity-modulated radiation therapy. J Appl Clin Med Phys 2012; 13 (2): 136145.CrossRefGoogle ScholarPubMed
Chen, J, Chen, X, Huang, M et al. A fixed-jaw method to protect critical organs during intensity-modulated radiotherapy. Med Dosim 2014; 39 (4): 325329.CrossRefGoogle ScholarPubMed
Feng, Z, Wu, H, Zhang, Y et al. Dosimetric comparison between jaw tracking and static jaw techniques in intensity-modulated radiotherapy. Radiat Oncol 2015; 10: 28.CrossRefGoogle ScholarPubMed
Shi, L, Lai, Y, Chen, S et al. Dosimetric superiority of IMRT with jaw tracking technique for whole esophagus and T-shaped field radiotherapy in advanced esophageal cancer. PLoS ONE 2018; 13 (9): e0202628.CrossRefGoogle ScholarPubMed
Kim, J, Park, J M, Park, S Y et al. Assessment of potential jaw-tracking advantage using control point sequences of VMAT planning. J Appl Clin Med Phys 2014; 15 (2): 160168.CrossRefGoogle ScholarPubMed
Snyder, K C, Wen, N, Huan, Y et al. Use of jaw tracking in intensity modulated and volumetric modulated arc radiation therapy for spine stereotactic radiosurgery. Pract Radiat Oncol 2015; 5: e155e162.CrossRefGoogle ScholarPubMed
Park, B D, Cho, B C, Kim, J H et al. Dosimetric impact of the jaw-tracking technique in volumetric modulated arc therapy. J Nucl Med Radiat Ther 2016; 7 (5): 301.Google Scholar
Wu, H, Jiang, F, Yue, H et al. A comparative study of identical VMAT plans with and without jaw tracking technique. J Appl Clin Med Phys 2016; 17 (5): 133141.CrossRefGoogle ScholarPubMed
Yao, S, Zhang, Y, Chen, T et al. Dosimetric comparison between jaw tracking and no jaw tracking in intensity-modulated radiation therapy. Technol Cancer Res Treat 2019; 18: 16.CrossRefGoogle ScholarPubMed
Mani, K R, Upadhayay, S, Das, K J M. Influence of jaw tracking in intensity-modulated and volumetric-modulated arc radiotherapy for head and neck cancers: a dosimetric study. Radiat Oncol J 2017; 35 (1): 90100.CrossRefGoogle ScholarPubMed
Thongsawad, S, Khamfongkhruea, C, Tannanonta, C. Dosimetric effect of jaw tracking in volumetric-modulated arc therapy. J Med Phys 2018; 43 (1): 5257.Google ScholarPubMed
Murakami, Y, Magome, T, Matsubayashi, F et al. Evaluation of organ-at-risk dose reduction with jaw tracking technique in flattening filter-free beams in lung stereotactic body radiation therapy. Phys Med 2019; 61: 7076.CrossRefGoogle ScholarPubMed
Pokhrel, D, Sanford, L, Halfman, M et al. Potential reduction of lung dose via VMAT with jaw tracking in the treatment of single-isocenter/two-lesion lung SBRT. J Appl Clin Med Phys 2019; 20 (5): 5563.CrossRefGoogle ScholarPubMed
Jeevanandam, P, Agnew, C E, Irvine, D M et al. Improvement of off-axis SABR plan verification results by using adapted dose reconstruction algorithms for the Octavius4D system. Med Phys 2018; 45 (4): 17381747.CrossRefGoogle Scholar
Liu, H, Ye, J, Kim, J J et al. Dosimetric comparison of two arc-based stereotactic body radiotherapy techniques for early-stage lung cancer. Med Dosim 2015; 40 (1): 7681.CrossRefGoogle ScholarPubMed
Shi, C, Chen, Y, Fang, D X et al. Application of modified dynamic conformal arc (MDCA) technique on liver stereotactic body radiation therapy (SBRT) planning following RTOG 0438 guideline. Med Dosim 2015; 40 (1): 2631.CrossRefGoogle ScholarPubMed
Shi, C, Tazi, A, Xiandong, D et al. Implementation and evaluation of modified dynamic conformal arc (MDCA) technique for lung SBRT patients following RTOG protocols. Med Dosim 2013; 38 (3): 287290.CrossRefGoogle ScholarPubMed
Smith, A, Kim, S, Serago, C et al. Use of flattening filter free photon beams for off-axis targets in conformal arc stereotactic body radiation therapy. World Congress on Medical Physics and Biomedical Engineering, June 7–12, Toronto, Canada. IFMBE Proceedings. Cham: Springer, 2015.CrossRefGoogle Scholar
Ross, C C, Kim, J J, Chen, Z J et al. A novel modified dynamic conformal arc technique for treatment of peripheral lung tumors using stereotactic body radiation therapy. Pract Radiat Oncol 2011; 1: 126134.CrossRefGoogle ScholarPubMed
Paddick, I. A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note. J Neurosurg 2000; 93 (suppl 3), 219222.CrossRefGoogle ScholarPubMed
Wessa, P. Free statistics software, office for research development and education, version 1.2.1, 2022. https://www.wessa.net/. Accessed on 23th September 2022.Google Scholar
Ng, J, Shuryak, I. Minimizing second cancer risk following radiotherapy: current perspectives. Cancer Manag Res 2015; 7: 111.Google ScholarPubMed
Ruben, J D, Davis, S, Evans, C et al. The effect of intensity-modulated radiotherapy on radiation-induced second malignancies. Int J Radiat Oncol Biol Phys 2008; 70 (5): 15301536.CrossRefGoogle ScholarPubMed
Tubiana, M. Can we reduce the incidence of second primary malignancies occurring after radiotherapy? A critical review. Radiother Oncol 2009; 91 (1): 415.CrossRefGoogle ScholarPubMed